QUBE COMPACT MONITOR 의 리콜

Health Canada (via FOI)에 따르면, 해당 리콜 는 Canada 에서 SPACELABS HEALTHCARE (CANADA) INC. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    142161
  • 사례 위험등급
    II
  • 사례 시작날짜
    2015-02-04
  • 사례 국가
  • 사례 출처
    HC
  • 비고 / 경고
    Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
  • 데이터 추가 비고
  • 원인
    Spacelabs has received multiple reports of xprezzon and qube monitors failing to boot up or returning to factory default configuration settings following power on or reset. reported issues include: -a failure to boot following power on or reset resulting in a failure in monitor operation. -loss of network connectivity which may result in i) an inability to display alarms and data from the affected monitor on central stations and other bedside monitors or ii) a failure to transmit parameter and alarms from the affected monitor to system printers and spacelabs' intesys clinical suite retrospective database (including the clinical events interface remote notification feature). -display of english language when previously configured for another language (the alarm messages displayed in the parameter waveform zone are not affected as these are stored on a command module and not on the monitor). -touchscreen failure where touching the screen does not produce the expected response. the parameter alarms and information are still present and correct once the monitor has completely started up. no one has been injured as a result of these issues. spacelabs believes this issue is caused by inadequate electrical contact on one of its printed circuit board assemblies.

Device

  • 모델명 / 제조번호(시리얼번호)
    Model Catalog: 91390 (Lot serial: > 10 lot numbers contact mfg); Model Catalog: 91393 (Lot serial: > 10 lot numbers contact mfg)
  • 제품 설명
    QUBE COMPACT MONITOR
  • Manufacturer

Manufacturer