Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
One report received of the actuation of ventilator alarm and buzzing sound after 16 hrs of patient oscillation in a hospital in germany. the reason was a failure in the bonding of the diaphragm chamber producing leakage.