Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Sysmex america inc. received complaints from sysmex installation specialists and customers that level 2 of the five levels of white blood cell (wbc) linearity of range check xii (rcxll) was recovering above the expected range on the poch-1 ooi xp-300 and kx-21 n and kx-21 analyzers (three-part differential analyzers). level 2 wbc recovery is expected to be approximately 0.5 +/- 0.5 x103/ul and recovery was above 1.0 x103/ul in some instances. when the recovery is above 1.0 the plotted line is not straight suggesting the analyzer is not linear through the entire wbc reportable range of 1.0 to 99.9 x103/ul for the three part differential analyzers. exclusion of the level 2 produces a linear graph.