데이터 추가 비고
원인
Siemens healthcare diagnostics has received several reports from customers of one or more electrolyte parameters (e.G. na+ ca++ k+ cl-) reporting a "-----?" error code when aqc qc or cvm is run on the rapidpoint 500 blood gas analyzer. when exhibited the tissue is usually seen during level 1 qc/aqc level 1 cvm and sometimes level 2 cvm testing. this issue has also been observed on a small number of patient samples. this issue impacts only rapidpoint 500 measurement cartridges p/ns 10491447 10491448 10491449 and 10844813. siemens has observed this issue with rapidpoint 500 measurement cartridges with serial numbers up to 2412705915 however not all these cartridges will exhibit this behavior. the issue does not impact rapidpoint 400 or rapidpoint 405 measurement cartridges.
Model Catalog: 10491447 (Lot serial: 2334500011 THROUGH 2412705915); Model Catalog: 10491448 (Lot serial: 2334500011 THROUGH 2412705915); Model Catalog: 10491449 (Lot serial: 2334500011 THROUGH 2412705915); Model Catalog: 10844813 (Lot serial: 2334500011 THROUGH 2412705915)
제품 설명
RAPIDPOINT 500 SYSTEM - LACTATE MEASUREMENT CARTRIDGES