Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens healthcare diagnostics has determined that when certain steps occur there is a potential for the first and/or last name of one patient to be printed with patient id and result data from a different patient even though those fields have been turned off for the system.
Model Catalog: 10492730 (Lot serial: 10697306); Model Catalog: 106444 (Lot serial: 10291507); Model Catalog: 10492730 (Lot serial: 10696855); Model Catalog: 10492730 (Lot serial: 10492730); Model Catalog: 05320281 (Lot serial: >10 CONTACT MANUFACTURER); Model Catalog: 129887 (Lot serial: >10 CONTACT MANUFACTURER); Model Catalog: 1265 (Lot serial: 10491395); Model Catalog: 1265 (Lot serial: 10321852); Model Catalog: 1265 (Lot serial: 10470366); Model Catalog: 1240 (Lot serial: 10491392); Model Catalog: 1240 (Lot serial: 10321840); Model Catalog: 106444 (Lot serial: 10339634); Model Catalog: 106444 (Lot serial: 10331381); Model Catalog: 106444 (Lot serial: 10328803); Model Catalog: 106444 (Lot serial: 10324081); Model Catalog: 106444 (Lot serial: 10322654); Model Catalog: 106444 (Lot serial: 10321239); Model Catalog: 106444 (Lot serial: 10318899); Model Catalog: 106444 (Lot serial: 10314585); Model Catalog: 10492730 (Lot serial: 10696857)
제품 설명
RAPIDPoint 500 Blood Gas Analyzer;RAPIDPoint 400 Blood Gas Analyzer;RAPIDPoint 405 Blood Gas Analyzer;RAPIDPoint 1265 Blood Gas Analyzer;RAPIDPoint 1260 Blood Gas Analyzer