Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
This notice concerns an issue found with the electron dose calculation in raystation 3.5 raystation 4.0 raystation 4.5 raystation 4.7 raystation 5 raystation 6 rayplan 1 (raystation 4.9) and rayplan 2 (raystation 6). when calculating electron monte carlo dose with a very large number of histories the dose calculation may be wrong.