원인
Returned product have shown that loss of pacing output could occur when pacemakers are exposed to strong electromagnetic fields such as modern anti-theft gates. no more than 0.6% are affected.
Model Catalog: 700 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 800 (Lot serial: ALL SERIAL NUMBERS); Model Catalog: 600 (Lot serial: ALL SERIAL NUMBERS)
“If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.