Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Stryker endoscopy has become aware that the sdc pro 2 picture/video capture device will not be able to assign the correct time/date stamp for saved files after 31 december 2009. on january 1st 2010 the system date for all sdc pro 2 units will re-start from october 1st 2001 and begin incrementing from that point onwards. this problem will not affect how pictures or videos are captured using sdc pro 2 during surgery. the system date will be incorrect and users will not be able to correct this issue.