원인
During testing shaver handpiece tray failed sterrad 100s half cycle testing and did not demonstrate a sterility assurance level of 10-6 for the tps shaver. it has been determined that the reprocessing guide (p10299 rev.A) is the non-conforming part of the shaver handpiece tray and the sterilization method with inadequate sterilization validation (sterrad 100s) needs to be removed as a sterilization option for the tps small joint shaver.