데이터 추가 비고
원인
Medtronic is issuing this notification following reports that the signia power handle may shut down unexpectedly during use. the signia power handle battery management system may incorrectly report the battery relative state of charge which can result in a display showing more battery life than is available. in some cases the power handle may have insufficient energy to complete the intended operation. use of the affected product may result in the need to use the manual retraction tool to remove the device from tissue.
“If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.