Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The centrifuge has potential of starting up unexpectedly when encountering an electrostatic fdischarge. the rotor may begin to spin when the bowl or spindle receives a shock of static electricity of approx 1.8kv which can happen in the typical user environment.