원인
Due to a pressure increase after the cylinder switch-over there are 2 possible issues identified:1. the cylinder status icon can change and erroneously indicate remaining gas in an empty no(nitric oxide) gas cylinder. 2. the alarm "no cylinder 1 missing (or empty)" or "no cylinder 2 missing (or empty)" can be deactivated. if the empty no cylinder is not replaced both no cylinders can run empty and cause a stop in no delivery. there have been no reports of patient injury as a result of this issue.