Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens has recently detected a rubber component with unfavorable ageing properties on the somatom emotion 16-slice configuration. the mechanical function of this rubber damper which is situated within the mounting assembly of the rotation motor may potentially degrade over time and may impact stability.