Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Baxter has identified a discrepancy between the loading/bolus default dose settings in the master drug library (mdl) version 8.0 and the values displayed on the pump during loading/bolus dose programming. when a drug's loading dose or bolus dose is configured in the mdl with a default starting time value in seconds the loading dose or bolus dose setup screen on the pump will display the time value rounded to the nearest integer in minutes. when this occurs the pump will administer drugs as configured. this discrepancy is only with the default time on the setup screen not with the amount or rate of drugs administered as configured in the drug library.