Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The sigma spectrum pump's keypad has nineteen physical hard and soft keys used to program and operate the pump. baxter has become aware that keys can become electrically shorted in field use and act as if they are depressed even without user input.