Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
On june 26 2014 one customer complaint was received for this issue (internal ref. # pr162312): a customer reported that a sample had discrepant results when using ssp unitray versus two other typing methods. ssp unitray typed the sample as dpb1*04:01dpb1*81:01. the correct assignment should be dpb1*04:01 dpb1*33:01. this is considered a mistype result.