Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
A limited quantity of the cmax hand controls were manufactured without loctite thread locker on the screws that attach the mounting clip to the hand control case. it is possible that screws without the loctite can work loose and the clip can become separated from the hand control. if both screws back out completely the internal components of the hand control and/or the table control board may become damaged. this could cause a loss of hand functionality.