원인
Livanova has determined that certain oxygenators containing a heat-exchanger with a polymer membrane separating the blood compartment from the water compartment when used with the cleaning an disinfection regimen described in the ifu for the heater cooler system 3t (ifu versions applicable since 2013) may result in diffusion of a small quantity of hydrogen peroxide from the water side to the blood side during a typical procedure in a quantity that exceeds the allowable limits determined by livanova.