Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
May experience calibration failure. the stratus cs instrument trigger the following error codes: slope and calcv error. calibrations that result in a failure prevents the d-dimer assay from being run on the stratus cs instrument until a successful calibration can be achieved.