원인
When the user imports two or more image series and changes the viewing direction for both image series there is a potential that the orientation of the second image series is not stored correctly by the system. this could result in two different discrepancies: 1) the automatic correlation algorithm may not achieve a good correlation result and 2) the second image series may be shown with the wrong orientation.