원인
Stryker endoscopy has initiated a voluntary product field correction for the instructions for use (ifu) for the suction irrigator reusable tips. the pre-vacuum steam sterilization (wrapped method) temperature in the instructions for use (ifu part # 1000-400-100 rev. k) incorrectly listed 121 - 123c instead of the correct 132 - 133c.