Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The problem is isolated to the blood order processing(bop) module "accession number" look-up mode of entry. once the "search" button is chosen for an accession number the system allows entry of a second accession number without first clearing the search results from the first accession number. this may cause the second accession number data to be combined with the first accession number data on the screen display when the "select" button is chosen. as a result if data is saved a patient other than the one expected may be updated.
Model Catalog: SQ1-BLB-SY (Lot serial: Version 6.0.1); Model Catalog: SQ1-BLD-SY (Lot serial: Version 6.0.1); Model Catalog: SQ1-BLD-SY (Lot serial: Version 6.0.2); Model Catalog: SQ1-BLB-SY (Lot serial: Version 6.0.2)