Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
During service suresight vision screeners were calibratedand the that the calibration activity performed may not have returned the device to working order. in its current state the device will most likely not be able to take a reading or it may be very difficult to get a reading. devices affected are limited to those serviced between july 17 2015 and aug 7 2015 at the welch allyn skaneateles falls location.