Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
ICU Medical, Inc. acquired from Pfizer in early 2017, Hospira Infusion Systems, the portion of Hospira dedicated to develop infusion pumps.
데이터 추가 비고
원인
Symbiq pole clamp assemblies used to repair the symbiq infuser have been identified to be missing the rubber stop pad. this pad reduces the potential for the symbiq infusion pump to slip down an iv pole. all serial numbers are potentially affected as they all have the potential for an affected pole clamp assembly to be installed on them during repair. the part number for the impacted pole clamp assembly is hsp3313-e12 the affected lot # is 481976.