Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
As a follow-up action to a previous corrective a new c4 reference range has been established. the new unicel/synchron c4 reference interval is 18 mg/dl to 55 mg/dl (s.I. units 180 mg/l to 550 mg/l). the previous reference range was 16 to 38 mg/dl (s.I. units 160-380 mg/l).