Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Siemens is notifying customers using syngo imaging studio / advanced studio version vb36c and older of the possibility of incorrect values for distance measurement. for measurement calculation syngo imaging vb36c and older version use the dicom attribute "imager pixel spacing" (tag (0018 1164))instead of "pixel spacing" (tag (0028 0030)). this may result in incorrect values for distance measurements. this happens if both dicom attributes "imager pixel spacing" and "pixel spacing" are set and have different values.