Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The recall was intiated as a result of findings that the product may not retain stability through out it's entire 36 month shelf life. the investigation has shown that there is a link between the age of product and the staining intensity and it was found the labelled 36 month shelf life was not appropriate.