Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Smiths medical has become aware of a small number of complaints of customers experiencing difficulty disconnecting accessories from the connectors of the affected tubes. in some cases the customer was unable to disconnect the accessory or excessive force lead to decannulation of the tube and an emergency tracheostomy tube change was required. if the accessory is correctly connected to the affected tubes then there will be no issue with disconnection.
Model Catalog: 67P040 (Lot serial: 1631477 through 1923406); Model Catalog: 67N030 (Lot serial: 1631477 through 1923406); Model Catalog: 67N035 (Lot serial: 1631477 through 1923406); Model Catalog: 67P055 (Lot serial: 1631477 through 1923406); Model Catalog: 67P050 (Lot serial: 1631477 through 1923406); Model Catalog: 67P045 (Lot serial: 1631477 through 1923406); Model Catalog: 67N025 (Lot serial: 1631477 through 1923406); Model Catalog: 67P035 (Lot serial: 1631477 through 1923406); Model Catalog: 67P030 (Lot serial: 1631477 through 1923406); Model Catalog: 67N040 (Lot serial: 1631477 through 1923406); Model Catalog: 65P055 (Lot serial: 1631477 through 1923406); Model Catalog: 60PFP40 (Lot serial: 1631477 through 1923406); Model Catalog: 65P045 (Lot serial: 1631477 through 1923406); Model Catalog: 65P040 (Lot serial: 1631477 through 1923406); Model Catalog: 65P035 (Lot serial: 1631477 through 1923406); Model Catalog: 65P030 (Lot serial: 1631477 through 1923406); Model Catalog: 65P025 (Lot serial: 1631477 thr