원인
An anomaly has been identified with the high - energy clinac gantry center throat cover where the cover may lean toward isocenter far enough to strike a patient. this issue is limited to center throat covers on high energy clinacs (h14 h27 h29) with v1.0 of the obi kv imaging system. these covers are associated with the obi air-cooled kv x-ray tube. other center throat covers use a different design.
Model Catalog: (Lot serial: H291166); Model Catalog: (Lot serial: H080220); Model Catalog: (Lot serial: H080286); Model Catalog: (Lot serial: H082013); Model Catalog: (Lot serial: H080059); Model Catalog: (Lot serial: H080217); Model Catalog: (Lot serial: H080213); Model Catalog: (Lot serial: H291172); Model Catalog: (Lot serial: H080194); Model Catalog: (Lot serial: H080268); Model Catalog: (Lot serial: H080060); Model Catalog: (Lot serial: H291194); Model Catalog: (Lot serial: H291021); Model Catalog: (Lot serial: H291178); Model Catalog: (Lot serial: H291016); Model Catalog: (Lot serial: H293307); Model Catalog: (Lot serial: H272370); Model Catalog: (Lot serial: H272407)