Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The manufacturer has determined through internal testing that the circuit that supplies backup power to the modules is not working. if there is a power interruption the 3-minute backup may not be operational and you may lose any changes you have made to the alarm settings. when used with model 91387 the hazard presents a risk for all modules except modules 90470 90478 90496 & 91496 all of which have their own internal backup and will not lose changes to the alarm settings for power interruptions less than 3 minutes.