Beckman coulter has confirmed low recovery values for reticulocytes may be generated from the unicel dxh 800 coulter cellular analysis system due to intermittent high electrical noise on the upper median angle light scatter (umal) produced by the volume conductivity scatter and multiple angle (vcsn) preamp board assembly part number (pn) b48187. the electrical noise produced by the vcsn preamp board assembly is greater than the instrument specification of 3.000 millivolt root mean square (mv rms).