데이터 추가 비고
원인
Beckman coulter has determined that a fitting installed in a select population of slidemaker stainer (sms) units may not completely seal after being installed. not all fittings installed may be affected. for those affected one may observe a slow dripping leak of the staining reagents (methanol alcohol and buffers) into the spill tray below designed for this purpose.