Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
The dxh sms (slidemaker stainer) system is generating an "invalid code" message (error code 75) when the lis transmits any field in the record that is not populated (nil). the lis will retransmit the test order to the analyzer and the dxh sms will reject the retransmission and send a message to the lis that the test order was already received/processed.