Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Beckman coulter has received and confirmed a report of a fire within the stainer module of a customer's dxh slidemaker stainer. the fire was fully contained within the stainer module. the event was preceded by a significant increase in the frequency of the automatic top-off of the module's methanol bath. beckman coulter is initiating this field action to reduce the potential for flammable liquids coming into contact with electrical components in the stainer module possibly causing a fire.