원인
Quality control material identified unacceptable vial to vial differences in dose response. while most of the observed toxo igg qc results continue to fall within the labeled acceptance ranges for the lots the heterogeneity of the qc 1 vials may cause qc 1 to return out of range results with greater frequency.
Model Catalog: A31590 (Lot serial: ); Model Catalog: A31590 (Lot serial: lot 09280QC 1 Vial lot 0a0020); Model Catalog: A31590 (Lot serial: lot 092821QC1 Vial lot 0a0021)
제품 설명
UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM - TOXO IGG ASSAY