원인
Des plaintes ont ete recues concernant un probleme de deversement de liquide dans la chambre de dilution en raison de la contre-pression accumulee dans ces appareils. si le tuyau d'evacuation du liquide interne ou externe est bouche ou pince la contre-pression s'accumule et peut causer le deversement de liquide dans la chambre de dilution. afin d'eviter une potentielle exposition aux dechets liquides il est important de suivre les instructions et recommandations fournie dans le manuel d'utilisation du manuel d'utilisation.