Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
If the user modifies the gantry start or stop angle or direction of a mlc acr teh leaf positions are closed without warning and the patient would receive almost zero dose.
Model Catalog: (Lot serial: H831257); Model Catalog: (Lot serial: H833652); Model Catalog: (Lot serial: H833504); Model Catalog: (Lot serial: H8T3829); Model Catalog: (Lot serial: H831127); Model Catalog: (Lot serial: H83826)