원인
Ortho clinical diagnostics (ocd) has identified an anomaly with vitros system software version 3.1. internal testing confirmed that when using calibrator barcode labels supplied with vitros chemistry products calibrator kit 2 calibration may not occur under specific circumstances. consignees are being advised not to use the calibrator barcode labels for vitros calibrator kit 2 until this anomaly is resolved in a future version of software.