Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Ortho clinical diagnostics has received several customer complaints of lower than expected albumin results for samples that were expected to generate results significantly above the measuring (reportable) range of 0.6 - 19.0 mg/dl. these samples may be negatively biased and give values within the reportable range without an initial dilution.