Ocd has received an increase in customer complaints related to slide dispense issues with some vitros chemistry products slides. this increase is due to the anti-backup platen (abp) failing to advance inside the vitros chemistry products slide cartridge body causing an increase in condition codes on the vitros system. the abp is a plastic platform that supports the slide stack inside the cartridge and is pushed along with the slides to the top of the cartridge during the slide dispense cycle. if the abp is jammed within the cartridge body a slide will not be properly positioned to dispense and the system will generate a condition code. this can lead to system downtime as the user must remove the cartridge and re-seat the abp in the slide cartridge or replace it with a new cartridge.