Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
While performing an internal investigation ocd determined that vitros chemistry products k+ slides coatings 0884 and 0885 contain a surfactant that was unintentionally added during the manufacturing process. the surfactant is used in several vitros slide products but it is not included in the product specifications for vitros k+ slides. ocd has conducted testing that indicates that vitros k+ slides made from coatings 0884 and 0885 are currently meeting their intended use and performance specifications. however at this time we are unable to determine the effect that the addition of the surfactant may have on long-term product performance.