원인
Ortho clinical diagnostics (ortho) identified the potential for biased results to be generated over the 10 day on-analyzer storage limit when using vitrosna+ slide cartridges warmed between 1.5 to 8 hours then placed on the analyzer. ortho has determined that cartridges require a minimum of 8 hours of warm up at room temperature to help ensure the slides have acceptable performance throughout the 10 day on-analyzer storage limit.