원인
Ortho confirmed that results obtained from vitros ipth reagent packs are positively biased (i.E. falsely elevated) compared to an alternative commercially available method. ortho observed a positive bias of 40% for samples with ipth concentrations <100pg/ml when testing with vitros ipth reagent packs in comparison to the roche elecsys pth test. ortho data demonstrated that this bias was consistent for all in-date lots.