Canadian data is current through March 2018. All of the data comes from Health Canada, except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Canada.
데이터 추가 비고
원인
Firm became aware of an event in which a 38mm pole broke due to excessive load. it was found that the draeger infinity acute care system was mounted on the optional 38mm pole in a way not validated by draeger and the maximum load of 13 kg was exceeded. if the weight of the equipment mounted on the 38mm pole exceeds 13kg it could potentially bend break or fall on a patient user or bystander.