원인
Zimmer is initiating a product recall for the listed zimmer trabecular metal reverse shoulder system glenosphere and base plate item numbers because it has received 90 complaints (complaint rate of 0.27%) related to the glenosphere disassociating from the base plate or difficulty assembling the glenosphere to the base plate. based on the investigation of these complaints zimmer determined that additional instructions specific to the assembly of the glenosphere to the base plate were necessary in order to emphasize important techniques.
“The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.