조치
DPC withdraws from the market lot 501 (per 30.06.03) the reagent called Immulite / HBsAg - ref.LKHB1 and lot 501 (per 30.06.03) of the reagent called: Immulite 2000 / HBsAg - ref. L2KHB2 due to a low cps (beats per second) activity of the calibrator, which may increase the incidence of false positive results. note: the systematic neutralization test for all positive results, as recommended in the technical notice, avoids the rendering of a false positive.