French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
데이터 추가 비고
조치
Lots 230242002 (per.25.02.01), 230250002 (per.05.03.01), 230255002 (per.10.03.01), 230269002 (per.24.03.01) of the reagent called: Stratus® CS / TestPak Troponin I Cardiac - CCTNI reference of the company DADE BEHRING SA are withdrawn from the market due to a high background noise resulting in an increase of the results with possibility, for normal samples, to exceed the threshold value (0.06ng / ml).