French data is current through early September 2018. All of the data comes from L’Agence nationale de sécurité du médicament et des produits de santé (ANSM), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and France.
데이터 추가 비고
조치
Lot 738182101 - lot of cartridge 000805-0 of the reagent called VIDAS CA 125 II - reference 30426 of the company BIOMÉRIEUX SA is withdrawn from the market following a change in the time of the performances of the reagent resulting in an overestimation of the results (currently order of 6U / ml max.) and a value out of control specifications.