All Pace Medical Inc temporary pacemakers and pacing system analysers (PSAs) 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 Pace Medical Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 연번
    MDA/2014/015
  • 날짜
    2014-05-14
  • 사례 출판 날짜
    2014-12-17
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
    These devices are currently not on the EU market. Their CE-marking has been suspended because of quality system failings.In April 2014, Pace Medical’s subcontractor Devices Ltd, was audited by its notified body, BSI, with MHRA present as the observer. The audit found significant quality system failings relating to processes for complaint handling, post-market surveillance and vigilance, which included failures to implement corrective actions agreed at previous audits. As a result of this, on 9 April, the notified body suspended the EC-certification covering all devices manufactured by Pace Medical, except the pacing cables. Servicing of devices already in the field is unaffected by the suspension of the EC certificate.We are aware that UK hospitals use a range of temporary pacing cables and heart wires with temporary pacemakers. However, Pace Medical only holds validation data to confirm compatibility of their own cables with their pacemakers. Hospitals should continue to make a risk-based decision on the use of alternative, reusable or disposable cables or heart wires, based on availability and clinical need.The manufacturer is working with its notified body to prepare a corrective action plan to address the identified issues. The notified body will need to complete follow-up audits of Pace Medical’s actions before the EC-certification can be restored. If the situation changes, MHRA will issue further information.
  • 원인
    (pace medical inc) ce-marking suspended due to quality system failings. (mda/2014/015).
  • 조치
    Be aware that as of 9 April 2014 this manufacturer is not able to sell these devices. However:any remaining Pace Medical devices purchased by distributors before this date can still be sold to UK hospitals servicing of Pace Medical devices can continue as before.Be aware that Pace Medical reusable temporary pacing cables, which have been validated by the manufacturer for compatibility with their temporary pacemakers, remain available for purchase.Be aware that the compatibility of alternative disposable or reusable cables with these pacemakers has not yet been fully validated by the manufacturer.Consult with the manufacturer or relevant distributors for further information on cable availability, where necessary.Consider the implications this has for the future provision of new devices within your hospital.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    All temporary pacemakers and pacing system analysers (PSAs) manufactured by Pace Medical Inc.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Manufacturer’s authorised representativeSteven Hanson or Catherine Rose Devices Ltd 28 Fiddlebridge Industrial Centre Lemsford Road Hatfield Hertfordshire AL10 0DE Tel:    01707 327 641 Fax:  01707 266 677Email:  pacemedicalpres@gmail.com
  • 제조사 모회사 (2017)
  • Source
    MHRA