Flexible endoscopes 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 Olympus 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 ID
    MDA/2007/034
  • 사례 연번
    CON2031041
  • 날짜
    2007-05-10
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
  • 원인
    The outer sheath and light guide tube of olympus flexible endoscopes may be damaged by direct exposure to ultraviolet (uv) light, as used in some endoscope storage and drying cabinets. keymed has informed the mhra of the increase in olympus flexible endoscopes being returned to them for repair due to damage to the outer sheath of the insertion tube and light guide tube. olympus believes that this damage is caused by prolonged exposure to uv light, as used in some endoscope storage and drying cabinets and causes significant deterioration in the outer coating of all olympus flexible endoscopes. this type of damage to the outer sheath of the flexible endoscope insertion tube and light guide tube may: cause trauma to the patient affect the functionality of the endoscope allow ingress of fluids into the endoscope inhibit effective decontamination of the endoscope by allowing micro-organisms to be retained in the damaged area of the endoscope reduce the number of endoscopes available for use in the unit whilst the affected endoscope is being repaired.
  • 조치
    Avoid placing Olympus flexible endoscopes in storage and drying cabinets where UV light shines directly onto the endoscope. If no other suitable storage or drying cabinet is available carry out a risk assessment for the continued use of these cabinets for storing and drying Olympus flexible endoscopes. Before purchasing a new endoscope ensure that it is compatible with local endoscope storage and drying cabinets. Be aware of MHRA advice given in MDA/2007/035.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Flexible endoscopes manufactured by Olympus, distributed in the UK by KeyMed Limited.